ULVAC DTC-41 USER MANUAL Pdf Download (2024)

ULVAC DTC-41 USER MANUAL Pdf Download (1)
  1. Manuals
  2. Brands
  3. Ulvac Manuals
  4. Water Pump
  5. DTC-41
  6. User manual

High vacuum diaphragm-type dry vacuum pump

Hide thumbs

1
Table Of Contents 2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
  • Contents
  • Table of Contents
  • Troubleshooting
  • Bookmarks

Table of Contents

  • Table of Contents

  • Before Using the Equipment

  • Using the Pump Safely

  • Safety Icons

  • Product Outline

  • Purpose of Use and Prohibitions

  • Specifications

  • Thermal Protection Relay

  • Dimensions

  • Installation and Storage

  • Cautions for Installation and Storage

  • Environmental Conditions for Installation, Storage, and Operation

  • Location

  • Checking Operation after Installation

  • Piping

  • Cautions for Safety in Use

  • Storage

  • Cautions for Operation

  • Operation of the Thermal Protection Relay

  • Starting in Cold Weather

  • Pump Performance

  • Pressure Achieved

  • Evacuation Rate

  • Power Requirements

  • Maintenance, Inspection, and Repair

  • Cautions for Maintenance, Inspection, and Repair

  • Maintenance

  • Regular Inspections

  • Replacing and Cleaning Consumables

  • Replacing Diaphragms

  • Replacing Inlet and Outlet Valves

  • Replacing O Rings

  • Replacing Bearings

  • Troubleshooting List

  • In Conclusion

  • Warranty

  • Usage Status Check Sheet

Advertisement

Quick Links

Download this manual

No.27300-2-01-11

User's Manual

High Vacuum Diaphragm-type Dry Vacuum Pump

Model: DTC-41 , 41K

DTC-41

DTC-41K

Request to Users

Please read this manual thoroughly to ensure safe and

effective use of the equipment.

Keep this manual in a safe place.

Due to periodic improvements in performance, the

equipment described in this manual is subject to

changes in dimensions and specifications without prior

notice.

ULVAC KIKO,Inc.

Table ofContents

NextPage
  • 1
  • 2
  • 3
  • 4
  • 5

Advertisement

Table of Contents

ULVAC DTC-41 USER MANUAL Pdf Download (2)

Related Manuals for Ulvac DTC-41

  • Water Pump Ulvac DTC-22 User Manual

    Diaphragm-type dry vacuum pump (27 pages)

  • Water Pump Ulvac DTU-20 Series Instruction Manual

    High vacuum diaphragm type dry vacuum pump (36 pages)

  • Water Pump Ulvac DTU-20 Instruction Manual

    Diaphragm dry vacuum pump (26 pages)

  • Water Pump Ulvac DTC-41C User Manual

    Diaphragm-type dry vacuum pump (28 pages)

  • Water Pump Ulvac DTC-22A User Manual

    Diaphragm-type dry vacuum pump (30 pages)

  • Water Pump Ulvac DTC-60 User Manual

    Diaphragm-type dry vacuum pump (26 pages)

  • Water Pump Ulvac DAP-9D-DC24 Instruction Manual

    Diaphragm-type dry vacuum pump (27 pages)

  • Water Pump Ulvac DOP-301SB Instruction Manual

    Oscillating piston dry vacuum pump (33 pages)

  • Water Pump Ulvac DOP-400SB Instruction Manual

    Oscillating piston dry vacuum pump (31 pages)

  • Water Pump Ulvac DOP-120SX User Manual

    Rocking piston type dry vacuum pump (31 pages)

  • Water Pump Ulvac DAT-50DA Instruction Manual

    Diaphragm-type dry vacuum pump (26 pages)

  • Water Pump Ulvac DA-41D Instruction Manual

    Diaphragm type dry vacuum pump (37 pages)

  • Water Pump Ulvac DA-121D Series Instruction Manual

    Diaphragm-type dry vacuum pump (32 pages)

  • Water Pump Ulvac DA-60D Instruction Manual

    Diaphragm dry vacuum pump (32 pages)

  • Water Pump Ulvac DAT-50DEP Instruction Manual

    Diaphragm dry vacuum pump (29 pages)

  • Water Pump Ulvac DA-30D Instruction Manual

    Diaphragm dry vacuum pump (28 pages)

Summary of Contents for Ulvac DTC-41

  • Page 1 No.27300-2-01-11 User’s Manual High Vacuum Diaphragm-type Dry Vacuum Pump Model: DTC-41 , 41K DTC-41 DTC-41K Request to Users Please read this manual thoroughly to ensure safe and effective use of the equipment. Keep this manual in a safe place. Due to periodic improvements in performance, the...
  • Page 2: Table Of Contents

    Contents Pages with a shaded background are those which contain items related to safety. Before Using the Equipment ....................01 Checks When Opening Packaging ..................02 Using the Pump Safely ......................02 • Safety icons ........................03 • Cautions for Safety in Use ....................04 1.
  • Page 3 Figures and Tables Fig.2.1 DTC-41 ........................2 Fig.2.2 DTC-41K ........................2 Fig.3.1 Example of Piping Used When Evacuating a Vessel ..........4 Table 1.1 Product Specifications ................... 1 Table 6.1 Consumables List ....................6 Table 6.2 Locations for Maintenance and Inspection ............. 7...
  • Page 4: Before Using The Equipment

    Before Using the Equipment Thank you for purchasing this product. Your custom is very much appreciated. This pump is designed solely for vacuum discharge, and may malfunction or cause accidents if not handled appropriately. Read the manual thoroughly, and pay due attention to inspections, maintenance, and safety.
  • Page 5: Checks When Opening Packaging

    Checks When Opening Packaging Check the following after opening the packaging. (1) Is the product as you requested? (2) Are the accessories and necessary parts included? Standard accessories ・User’s manual -------------------- x 1 ・Inlet and outlet caps (fitted to inlet and outlet) -------------------- x 2 ・Power plug adapter (attached to power cord) -------------------- x 1...
  • Page 6: Safety Icons

    ・Safety icons The meanings of the safety icons are as follows. Danger Incorrect handling of the equipment is very likely to result in death or serious injury to the operator. Warning Incorrect handling of the equipment may result in death or serious injury to the operator. Caution Incorrect handling of the equipment may result in light or medium injuries to the operator or damage to the equipment.
  • Page 7 Cautions for Safety in Use Danger Applications (1) This pump is not designed to be explosion-proof, and should therefore not be used to discharge explosive gases. (2) In addition to discharge of gas via the outlet, gas may also leak from other parts of the pump, and it should therefore not be used with toxic gases.
  • Page 8 Warning Operation (14) This pump is not designed to be explosion-proof. When using the pump, ensure that there are no inflammable materials such as solvents, or explosive gases, in the vicinity. Use under such conditions may result in injury or fire. (15) Inserting fingers or objects into the motor inlet may result in electric shock, injury, or fire.
  • Page 9 Caution Maintenance and Repair Dispose in accordance with legislation for disposal and cleaning of waste products, handle as industrial waste, and do not incinerate. Toxic fluorine gas is generated by incineration of fluorine-based plastics. (10) If the pump ceases operation, turn power OFF (set switch to O) immediately to prevent accidents, remove the power cord from the wall outlet, and contact your dealer or the manufacturer for inspection and repair.
  • Page 10 Note Installation Ingestion of liquids into the pump will result in damage and prevent proper operation. Ingestion of rubbish and dust in the air entering the pump will interfere with its proper function. If the air is likely to contain rubbish or dust, a filter should be fitted to the inlet to protect the pump. Ducting should always be fitted to the pump outlet if toxic corrosive gases, or steam, enters the pump.
  • Page 11: Product Outline

    (5) Ensure that the gas entering the pump does not contain rubbish, dust, or water (except steam). (6) Do not operate the pump for long periods at near-atmospheric pressure. 1.2 Specifications Table 1.1 Product Specifications Model DTC-41 DTC-41K 50Hz 40L/min Discharge rate 60Hz 46L/min 1.0×10...
  • Page 12: Thermal Protection Relay

    1.3 Thermal Protection Relay 1) This pump is fitted with an automatic reset thermal protection relay for overload protection. This device shuts off the motor power supply circuit automatically to prevent burn-out if the motor temperature rises due to a pump fault which prevents rotation, or if load becomes excessive.
  • Page 13: Installation And Storage

    3. Installation and Storage 3.1 Cautions for Installation and Storage See Warning Warning (1)(2)(3)(5)(6)(7)(8)(9)(10)(11)(12)(13), P04 Caution See Caution (1)(2)(3), P05 Note See Note (1)(2)(7), P06 3.2 Environmental Conditions for Installation, Storage, and Operation The fine clearances used in this pump require that the following conditions be satisfied during storage, installation, and operation.
  • Page 14: Cautions For Safety In Use

    Shut-off valve Vessel Filter Leak valve Vacuum pump Fig.3.1 Example of Piping Used When Evacuating a Vessel 3.6 Storage Turn the switch OFF (set to O), remove the power plug from the outlet, place the rubber caps over the inlet and outlet, and store the pump in an area of low humidity. 4.
  • Page 15: Pump Performance

    5. Pump Performance 5.1 Pressure Achieved The term “pressure achieved” as employed in the catalogue and in this manual is defined as “the minimum pressure obtained by the pump without introduction of gas from the pump inlet (ie the no-load condition)”. Note that the indicator values for pressure may differ between types of vacuum gauges.
  • Page 16: Maintenance, Inspection, And Repair

    6. Maintenance, Inspection, and Repair 6.1. Cautions for Maintenance, Inspection, and Repair Danger See Danger (3), P04 Warning See Warning (4)(17)(18), P04,05 Caution See Caution (9)(10)(11), P06 Note See Note (12), P07 Maintenance and repair by the customer’s repair technician is limited to the following procedures. Do not undertake other repairs, or make modifications other than the standard options supplied by the manufacturer.
  • Page 17: Replacing And Cleaning Consumables

    <Replacement Guide> Replace or clean components if performance is reduced or the following symptoms become apparent. Diaphragms : Replace if PTFE components are worn or peeling, or if rubber components are deformed, hard or cracked. Inlet and outlet valves : Replace if deformed, hard, or cracked. O rings : Replace if hard, cracked, or stretched.
  • Page 18: Replacing Diaphragms

    Caution: To prevent injury, always wear gloves when fitting or removing diaphragms. Use tools No.1, 2, 3, 4, 5, 6 and 7. 1-1). DTC-41 (1) Remove the inlet and outlet pipes, and lay the pump horizontally on a soft Photo.1 cloth (Photo.1).
  • Page 19 (6) When one of the diaphragms is pushed downwards (to Bottom dead Diaphragm Top dead bottom dead center), the other diaphragm moves upwards (to center center top dead center). The edge of the diaphragm may then be held with the fingers (Photo.6). Photo.6 (7) While holding the edge of the diaphragm, turn it in the anti-clockwise Photo.7...
  • Page 20: Replacing Inlet And Outlet Valves

    (14)Lay the pump horizontally and fix the front panel in place with the truss head screws (M4 x 10). (15)Stand the pump upright and fit the inlet and outlet pipes. 1 -2). DTC-41K Replace diaphragms in accordance with '1-1). DTC-41 ' steps (4)~(12). 2) Replacing Inlet and Outlet Valves (simultaneous replacement of both diaphragms is recommended) Use tools No.1, 2, 3, 4, 6 and 7.
  • Page 21: Replacing O Rings

    Caution: Repeatedly tighten equally in diagonally opposite pairs to a torque of 0.5N・m. (7) Assemble the pump in accordance with ‘1)Replacing diaphragms’ step(1)~(15). 2-2) DTC-41K Replace inlet and outlet valves in accordance with '2-1). DTC-41 ' steps (2)~(7). 3) Replacing O rings (simultaneous replacement of O rings and diaphragms is recommended) Use tools No.1, 2, 3, 4, 5, 6 and 7.
  • Page 22: Troubleshooting List

    6.5 Troubleshooting List Table 6.3 Troubleshooting List Problem Causes Solutions Reference (1) Not connected to power supply. (1) Connect power supply. (2) Switch is OFF. (2) Set switch to I. (3) Ensure that voltage variation is (3) Problem with power supply voltage. within +/-10%.
  • Page 23: In Conclusion

    7. In Conclusion Please contact the manufacturer’s sales division if you have any questions. Warranty (1) The warranty for this pump (this equipment) extends for a period of one year from the date of shipment. (2) Any malfunctions or defects which occur under normal usage conditions during the warranty period will be repaired free of charge.
  • Page 24: Usage Status Check Sheet

    Usage Status Check Sheet (for use in Instruction Manual) * For the purpose of safety control of repair personnel, fill in within the heavy line frame and attach the sheet to the item of which repair is requested. * In case this sheet were not attached or filled in, your request of repair and service may not be accepted.
  • Page 25 株式会社アルバック 規格品事業部 東日本営業部 〒253-8543 神奈川県茅ヶ崎市萩園2500 TEL:0467-89-2416 株式会社アルバック 規格品事業部 西日本営業部 〒532-0003 大阪府大阪市淀川区宮原3-3-31 上村ニッセイビル5F TEL:06-6397-2286 ULVAC KIKO,Inc. https://ulvac-kiko.com/en Please contact us for products, Service Base or other Inquiries from here. https://showcase.ulvac.co.jp/en/ ULVAC,Inc. Components Division 2500 Hagisono, Chigasaki, Kanagawa, 253-8543, Japan TEL:+81-467-89-2261...

This manual is also suitable for:

Dtc-41k

ULVAC DTC-41 USER MANUAL Pdf Download (2024)
Top Articles
Latest Posts
Recommended Articles
Article information

Author: Lilliana Bartoletti

Last Updated:

Views: 5251

Rating: 4.2 / 5 (53 voted)

Reviews: 92% of readers found this page helpful

Author information

Name: Lilliana Bartoletti

Birthday: 1999-11-18

Address: 58866 Tricia Spurs, North Melvinberg, HI 91346-3774

Phone: +50616620367928

Job: Real-Estate Liaison

Hobby: Graffiti, Astronomy, Handball, Magic, Origami, Fashion, Foreign language learning

Introduction: My name is Lilliana Bartoletti, I am a adventurous, pleasant, shiny, beautiful, handsome, zealous, tasty person who loves writing and wants to share my knowledge and understanding with you.